XCM BIOLOGIC TISSUE MATRIX
Report
- Report Number
- 2530154-2013-00014
- Event Type
- Injury
- Date Received
- January 10, 2014
- Date of Event
- December 3, 2013
- Report Date
- January 10, 2013
- Manufacturer
- DSM BIOMEDICAL
- Product Code
- FTM
- PMA / PMN Number
- K091499
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
METHOD: EVALUATION OF THIS EVENT IS LIMITED TO MEDICAL ASSESSMENT OF THE COMPLAINT NARRATIVE. THE EXPLANTED DEVICE IS NOT AVAILABLE FOR RETURN TO THE COMPANY. TO DATE, THE COMPANY HAS MADE MULTIPLE ATTEMPTS TO OBTAIN THE DEVICE LOT NUMBER, BUT IT HAS NOT BEEN PROVIDED. THEREFORE, LOT HISTORY RECORDS COULD NOT BE REVIEWED. IF THE LOT NUMBER BECOMES AVAILABLE AFTER THE SUBMISSION OF THIS INITIAL REPORT, THE LOT HISTORY RECORDS WILL BE EVALUATED AND THE RESULTS WILL BE REPORTED TO FDA IN A F/U REPORT. CONCLUSION: EARLY POSTOPERATIVE BOWEL OBSTRUCTION AFTER ABDOMINAL SURGERY CAN OCCUR WITHOUT MESH, OR WITH ANY MESH PRODUCT, SYNTHETIC OR BIOLOGIC, PERMANENT OR ABSORBABLE. THE BOWEL OBSTRUCTION IS MOST LIKELY RELATED TO PATIENT FACTORS.
THIS PATIENT PRESENTED WITH A RECURRENT, LARGE VENTRAL HERNIA WITHOUT BOWEL OBSTRUCTION. THE NEW HERNIA WAS LOCATED ABOVE THE MARLEX MESH IN THE MIDLINE WITH SYMPTOMS AND OVERALL LAXITY OF ENTIRE ABDOMINAL WALL. SURGERY INCLUDED EXCISION OF THE MARLEX MESH AND INLAY OF XCM BIOLOGIC UNDER THE RECTUS MUSCLES WITH INTERRUPTED SUTURES, FOLLOWED BY ADVANCEMENT OF THE RECTUS MUSCLES TO MIDLINE AND OVERLAY MESH (SANDWICH TECHNIQUE). THREE DRAINS WERE PLACED (ONE IN BETWEEN MUSCLES, TWO IN SUBCUTANEOUS TISSUES. THE PATIENT WAS DISCHARGED 5 DAYS LATER AND WAS DOING WELL. SHE HAD A BOWEL MOVEMENT AT HOME. THREE DAYS LATER, SHE HAD DISTENSION AND NAUSEA WITH A LOW GRADE TEMPERATURE. SHE WAS ADMITTED AND HAD AN OROGASTRIC TUBE PLACED WITH SEVERAL LITERS OF BILIOUS DRAINAGE. SHE WAS TAKEN BACK TO THE OPERATING ROOM AND FOUND TO HAVE ADHESION OF THE ENTIRE BOWEL TO XCM GRAFT. NO PERFORATION WAS OBSERVED. THERE WERE NO SUTURES WHICH HAD CAUGHT THE BOWEL. THE ADHESION CAUSED A POINT OF FIXATION WHERE THE SMALL BOWEL OBSTRUCTED (MECHANICAL OBSTRUCTION). THE PATIENT WAS STARTED ON IV ANTIBIOTIC. XCM WAS EXPLANTED AND THE MUSCLES WERE REPAIRED. THE PATIENT WAS DISCHARGED 5 DAYS LATER AND IS DOING FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19448 | XCM BIOLOGIC TISSUE MATRIX | OKY - MESH, SURGICAL, COLLAGEN, LG ABDOM | FTM | DSM BIOMEDICAL | 30010-32 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Hospitalization| L| R |