FDA Adverse Event Other Summary report: N

OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 3593333 · Received January 8, 2014

Report

Report Number
3008011247-2014-00004
Event Type
Other
Date Received
January 8, 2014
Date of Event
December 13, 2013
Report Date
January 20, 2015
Manufacturer
TRIVASCULAR, INC.
Product Code
MIH
PMA / PMN Number
P120006
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

BASED ON FOLLOW-UP INFORMATION RECEIVED FROM THE DISTRIBUTOR ON JANUARY 20, 2015, THE PATIENT RETURNED FOR A RE-INTERVENTION ON (B)(6) 2013 WHERE AN AORTIC CUFF WAS PLACED WITHIN THE AORTIC BODY SEALING RINGS AND SUCCESSFULLY RESOLVED THE TYPE IA ENDOLEAK. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL IN (B)(6) 2014 WITH NO ADDITIONAL PATIENT SEQUELAE REPORTED. FOLLOWING THE RE-INTERVENTION PROCEDURE IN (B)(6) 2013, THERE HAVE BEEN THREE (3) FOLLOW-UP CT'S PERFORMED IN 3-4 MONTHS INTERVALS, ALL OF WHICH CONTINUE TO SHOW EXCLUSION OF THE ANEURYSM.

Description of Event or Problem · 1

THE PT UNDERWENT ENDOVASCULAR REPAIR OF (B)(4) WITH THE OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM ON (B)(6) 2013. THE STENT GRAFT SYSTEM WAS IMPLANTED AND THE AORTIC BODY AND ILIAC LIMBS WERE BALLOONED. UPON FINAL ANGIOGRAM, THE AORTIC BODY WAS OBSERVED TO BE APPROX 1CM DISTAL TO THE INTENDED LANDING SITE, RESULTING IN A TYPE IA ENDOLEAK. BASED ON COMMUNICATION FROM THE CASE SUPPORTER, THE STIFF GUIDEWIRE WAS NOT RETRACTED DURING FILL WHICH CAN LEAD TO POOR APPOSITION OF THE SEAL RINGS TO THE VESSEL WALL. THE PHYSICIAN BELIEVED THE ENDOLEAK WAS SMALL AND THAT IT MIGHT RESOLVE ITSELF. A FOLLOW-UP CT WAS PLANED FOR (B)(6) 2013 TO DETERMINE IF A RE-INTERVENTION WOULD BE REQUIRED. AS OF THE DATE OF THIS REPORT, THERE HAS BEEN NO CONFIRMATION THAT A CT SCAN OR REINTERVENTION WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
14509 OVATION PRIME ABDOMINAL STENT GRAFT SYSTEM ENDOVASCULAR GRAFT, AORTIC ANEURYSM MIH TRIVASCULAR, INC. TV-AB3480-D FS051512-16

Patients

Seq Age Sex Outcome Treatment
1 83 YR Other| R GORE AORTIC CUFF