FDA Adverse Event Malfunction Summary report: N

ARCHITECT IPTH

MDR report key: 3591741 · Received January 24, 2014

Report

Report Number
3002809144-2014-00012
Event Type
Malfunction
Date Received
January 24, 2014
Date of Event
January 2, 2013
Report Date
January 9, 2014
Manufacturer
ABBOTT GERMANY
Product Code
CEW
PMA / PMN Number
K063232
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER PROVIDED RESULTS FROM (B)(4) PATIENT SAMPLES TESTED WITH THE ARCHITECT METHOD VERSES THE ROCHE METHOD. THE RESULTS ARE LISTED BELOW: COLLECTED-DATE: (B)(6) 2013, ARCHITECT: 64, ROCHE: 43; (B)(6) 2013, 314, 180; (B)(6) 2014, 116, 56; (B)(6) 2013, 119, 43; (B)(6) 2013, 83, 7; 2013, 684, 300; 2013, 30, 14; 2013, 17,8 89; (TEST RESULT). (B)(4). PRODUCT EVALUATION IS IN PROCESS AND THE EVALUATION RESULTS WILL BE SUBMITTED IN A FOLLOW-UP REPORT.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT PATIENT RESULTS FOR THE ARCHITECT IPTH ASSAY ARE BEING QUESTIONED BY PHYSICIANS SINCE THEY ARE HIGHER THAN EXPECTED. SEVERAL PATIENT SAMPLES WERE SPLIT BY THE DOCTORS AND SENT OUT TO BE TESTED BY A DIFFERENT METHOD. THE RESULTS WERE LESS THAN THE ARCHITECT RESULTS BY APPROXIMATELY 50% SHIFT. ONE PATIENT SAMPLE GENERATED A RESULT OF 119 (UNIT OF MEASURE WAS NOT PROVIDED) BY THE ARCHITECT AND A RESULT OF 43 WAS GENERATED BY THE ROCHET METHOD. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56449 ARCHITECT IPTH CEW ABBOTT GERMANY

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT I2K 03M74-01 SN (B)(4)