FDA Adverse Event
Injury
Summary report: N
THE INFINITY TRANSILLUMINATOR
MDR report key: 358788
·
Received October 30, 2001
Report
- Report Number
- 358788
- Event Type
- Injury
- Date Received
- October 30, 2001
- Date of Event
- October 15, 2001
- Report Date
- October 30, 2001
- Manufacturer
- FIBEROPTIC MEDICAL PRODUCTS, INC
- Product Code
- MQW
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
INFANT SUFFERED TWO, SECOND-DEGREE BURNS FOLLOWING TRANSILLUMINATION TO FACILITATE PUNCTURE OF THE LEFT ULNAR WRIST ARTERY. THE INFRARED BLOCKING FILTER WAS FOUND TO BE DISLODGED, RESULTING IN INCREASED LIGHT INTENSITY AND INCREASED TEMPERATURE TO THE SKIN SURFACE OF THE BACK OF THE WRISTS. THE INJURY CONSISTED OF TWO, SECOND-DEGREE BURNS OF 1/2CM EACH TO THE ANTERIOR SIDE OF THE WRISTS. EACH SECOND-DEGREE BURN RESULTED IN A RAISED, FLUID-FILLED VESICLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 48647 | THE INFINITY TRANSILLUMINATOR | PORTABLE HIGH-POWERED FIBEROPTIC EXAMINATION LIGHT | MQW | FIBEROPTIC MEDICAL PRODUCTS, INC | IT-101 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3.5 MO | Other| R |