FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3586663 · Received January 22, 2014

Report

Report Number
2531779-2014-02250
Event Type
Malfunction
Date Received
January 22, 2014
Report Date
January 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD WAS DAMAGED. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. THE KEYPAD COVER HAS A HOLE BELOW THE OK BUTTON AND IS TORN ACROSS THE OK BUTTON. THE BUTTON CONTACT IS EXPOSED. TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE CONTRAST BUTTON. MULTIPLE BUTTON PRESSES WITH INCREASED FORCE APPLIED ARE REQUIRED BEFORE THE CONTRAST BUTTON ENGAGES. THE UP, DOWN AND OK BUTTONS RESPOND TO USER INPUT. THERE WAS CONTAMINATION UNDER THE CONTACTS OF ALL BUTTONS. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE BUTTONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
53071 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 13 YR