AU600 CLINICAL CHEMISTRY ANALYZER
Report
- Report Number
- 9612296-2014-00011
- Event Type
- Malfunction
- Date Received
- January 21, 2014
- Date of Event
- December 30, 2013
- Report Date
- December 30, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JGS
- PMA / PMN Number
- K961274
- Removal / Correction Number
- Z-0152-0153-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER PROVIDED CALIBRATION AND QUALITY CONTROL DATA COLLECTED PRIOR TO THE EVENT. THE DATA WAS REVIEWED: CALIBRATION SLOPES GENERATED WERE LOW BUT WITHIN SPECIFIED RANGE (-38 TO -68) PRIOR TO THE EVENT ON (B)(6) 2013. QUALITY CONTROL DATA GENERATED WAS ALSO WITHIN TARGETS PRIOR TO THE EVENT. ON (B)(4) 2013, THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND DETERMINED THE REFERENCE VALVE WAS ALLOWING BUBBLES INTO THE FLOW CELL INTERMITTENTLY. THE FSE REPLACED THE REFERENCE VALVE AND PRIMED THE SYSTEM A NUMBER OF TIMES. NO FURTHER BUBBLES WERE OBSERVED. THE LOW CALIBRATION SLOPES ALSO IMPROVED AS A RESULT OF THE REFERENCE VALVE CHANGE. IDENTIFICATION OF FAILURE MODE: THE ERRONEOUS SODIUM RESULTS WERE MOST LIKELY CAUSED BY AN ISE REFERENCE VALVE MALFUNCTION. AS PER CUSTOMER LETTER FSN - URGENT FIELD SAFETY NOTICE 21338-1 : INACCURATE TEST RESULTS MAY BE GENERATED DUE TO THE PRESENCE OF AIR BUBBLES IN THE ISE REF SOLUTION TUBING. THE MAGNITUDE OF ERROR CANNOT BE PREDICTED. AFFECTED RESULTS MAY BE CLINICALLY SIGNIFICANT (HIGH OR LOW) AND MAY IMPACT PATIENT DIAGNOSIS AND/OR TREATMENT.
A CUSTOMER REPORTED ERRONEOUS LOW SODIUM TEST RESULTS WERE OBTAINED FOR APPROXIMATELY 30 PATIENT SAMPLES WHEN USING THE AU680 CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER STATED THAT THESE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY; HOWEVER, THERE WERE NO REPORTS OF ADVERSE AFFECTS TO PATIENT TREATMENT. THE TEST RESULTS WERE DEEMED TO BE ERRONEOUSLY LOW AFTER BEING QUESTIONED BY A PHYSICIAN AND THE SAMPLES WERE RETESTED. THERE WAS NO DEATH OR INJURY REPORTED FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47067 | AU600 CLINICAL CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JGS | BECKMAN COULTER | AU681-10E | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |