FDA Adverse Event Malfunction Summary report: N

MEDI-THERM

MDR report key: 35846 · Received August 21, 1996

Report

Report Number
MW1009778
Event Type
Malfunction
Date Received
August 21, 1996
Date of Event
July 15, 1996
Report Date
August 21, 1996
Manufacturer
GAYMAR INDUSTRIES, INC.
Product Code
DWJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

SPILLAGE OR CONDENSATION DRIPPED ONTO MAIN POWER FEED LUG ON BACK OF MAIN POWER SWITCH. THIS EITHER DIRECTLY STARTED OVERHEAT3ING THAT CONNECTION OR OVER TIME CORRODED CASUING IMPEDENCE OF CONNECTION TO INCREASE AND THEREBY CAUSING OVERHEATING OF CONNECTION. CO NOW SENDS SPLASH COVER WITH SWITCH, BUT MAY NOT HELP IF THERE IS INTERNAL CONDENSATION. THIS IS SECOND POWER SWITCH REPLACED WITH THIS FAULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDI-THERM HYPOTHERMIA BLANKET DWJ GAYMAR INDUSTRIES, INC. MTA *

Patients

Seq Age Sex Outcome Treatment
1 *