FDA Adverse Event
Malfunction
Summary report: N
MEDI-THERM
MDR report key: 35846
·
Received August 21, 1996
Report
- Report Number
- MW1009778
- Event Type
- Malfunction
- Date Received
- August 21, 1996
- Date of Event
- July 15, 1996
- Report Date
- August 21, 1996
- Manufacturer
- GAYMAR INDUSTRIES, INC.
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
SPILLAGE OR CONDENSATION DRIPPED ONTO MAIN POWER FEED LUG ON BACK OF MAIN POWER SWITCH. THIS EITHER DIRECTLY STARTED OVERHEAT3ING THAT CONNECTION OR OVER TIME CORRODED CASUING IMPEDENCE OF CONNECTION TO INCREASE AND THEREBY CAUSING OVERHEATING OF CONNECTION. CO NOW SENDS SPLASH COVER WITH SWITCH, BUT MAY NOT HELP IF THERE IS INTERNAL CONDENSATION. THIS IS SECOND POWER SWITCH REPLACED WITH THIS FAULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDI-THERM | HYPOTHERMIA BLANKET | DWJ | GAYMAR INDUSTRIES, INC. | MTA | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |