FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3580775 · Received January 17, 2014

Report

Report Number
2531779-2014-01833
Event Type
Malfunction
Date Received
January 17, 2014
Report Date
December 23, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/23/2013 WITH THE FOLLOWING FINDINGS: THE BLACK BOX REVIEW SHOWS TWO ¿CS088¿ WATCHDOG ALARMS ON THE REPORTED EVENT PERIOD. THE PUMP POWERS ON AND DISPLAYS VERIFY SCREEN. THE PUMP WAS PRIMED CORRECTLY, ¿CS088¿ ALARM OCCURRED DURING THE 24 HOUR DURATION TEST WITH CONSISTENT BUZZING SOUND. THE PUMP¿S COVER WAS REMOVED; INSPECTION SHOWS AN INTERMITTENT CONDITION TO THE PCB ILLUSTRATED WITH AN OPEN TRACE BETWEEN THE ¿U38-WATCHDOG CHIP¿ AND THE¿ MASTER MICROPROCESSOR. (B)(4).

Description of Event or Problem · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/23/2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT AN INTERMITTENT CONDITION TO THE PCB ILLUSTRATED WITH AN OPEN TRACE. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
44323 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR