NAVIOS? FLOW CYTOMETER SYSTEM
Report
- Report Number
- 1061932-2014-00110
- Event Type
- Malfunction
- Date Received
- January 16, 2014
- Date of Event
- December 16, 2013
- Report Date
- December 17, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- OYE
- PMA / PMN Number
- K130373
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE ON (B)(4) 2013. THE FSE EVALUATED THE GALLIOS FLOW CYTOMETER AND CONFIRMED THAT THE VORTEXER HAD FAILED. THE FSE RETURNED TO THE CUSTOMER'S SITE ON (B)(4) 2013 TO REPLACE THE MULTI CAROUSEL LOADER (MCL) LIFTER MOTOR ASSEMBLY ON THE GALLIOS FLOW CYTOMETER, MAKING THE UNIT OPERATIONAL. FAILURE MODE OF THE EVENT IS ATTRIBUTED TO THE VORTEXER MIXER MOTOR. (B)(4).
THE CUSTOMER REPORTED THAT THE VORTEXER TO THE MULTI CAROUSEL LOADER (MCL) SIDE OF THE GALLIOS FLOW CYTOMETER IS NOT WORKING. THE GALLIOS FLOW CYTOMETER IS A NOT AN IN-VITRO DIAGNOSTIC (IVD) INSTRUMENT AND IS FOR RESEARCH USE ONLY. THIS REPORT REFERENCES THE NAVIOS FLOW CYTOMETER SYSTEM WHICH IS AN IVD INSTRUMENT AND IS SIMILAR IN DESIGN AND FUNCTION TO THE GALLIOS FLOW CYTOMETER. THE VORTEXER IS A COMMON PART ON BOTH INSTRUMENTS AND IS PART OF THE MCL LIFTER MOTOR ASSEMBLY. ITS PURPOSE IS TO LIFT THE SAMPLE TUBE FROM THE CAROUSEL AND VORTEX (MIX) THE SAMPLE JUST PRIOR TO ASPIRATION BY THE INSTRUMENT. NO ERRONEOUS RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. GALLIOS FLOW CYTOMETER PART NUMBER A94304 WITH SERIAL NUMBER (B)(4) WAS INVOLVED IN THIS EVENT. THIS GALLIOS' INSTRUMENT DATE OF MANUFACTURE IS MARCH 13, 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 42300 | NAVIOS? FLOW CYTOMETER SYSTEM | FLOW CYTOMETRIC REAGENTS AND ACCESSORIES. | OYE | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |