FDA Adverse Event Malfunction Summary report: N

SOFT IMPACT MALLET W/ EASY GRIP HANDLES

MDR report key: 3572125 · Received January 2, 2014

Report

Report Number
1833053-2013-00009
Event Type
Malfunction
Date Received
January 2, 2014
Date of Event
October 4, 2013
Report Date
December 26, 2013
Manufacturer
MEDTORQUE INC.
Product Code
HXL
Product Problem
Yes
Report Source
Distributor report
Reporter Location
AK, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

WHILE IN USE DURING SURGERY, THE DELRIN HEAD OF THE MALLET BROKE OFF. THE DELRIN HEAD WAS FOUND AND THE DEVICE SEQUESTERED. NO INJURIES WERE SUSTAINED BY THE PATIENT. THE USER FACILITY WAS INFORMED THAT THE PART NUMBER HAD UNDERGONE A DESIGN CHANGE TO ALLOW ENHANCED IMMOBILITY UPON IMPACT. TO ACCOMPLISH THIS, THE RE-DESIGN INCORPORATED A METAL INSERT WITHIN THE THREADS OF THE HEADS TO CREATE A MORE TIGHT GRIP. NOTE THAT THE DESIGN CHANGE WAS ALREADY IN PROCESS WHEN THIS EVENT OCCURRED. THE USER FACILITY WAS ISSUED TWO LOANER DEVICES TO USE UNTIL THE DESIGN CHANGE PROCESS WAS COMPLETE. AFTER NEW DEVICES WERE MANUFACTURED, REPLACEMENTS WILL BE SENT TO THE CUSTOMER. A FOLLOW UP MESSAGE WAS LEFT REGARDING THE REPLACEMENT DEVICES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2649 SOFT IMPACT MALLET W/ EASY GRIP HANDLES HXL MEDTORQUE INC. 7832 1012MT

Patients

Seq Age Sex Outcome Treatment
1