OLYMPUS KEYMED
Report
- Report Number
- 9611174-2013-00012
- Event Type
- Malfunction
- Date Received
- November 8, 2013
- Date of Event
- October 15, 2013
- Report Date
- October 15, 2013
- Manufacturer
- KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT), LTD.
- Product Code
- FEM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
INITIAL INVESTIGATION BY OLYMPUS (B)(4) FIELD SERVICE ENGINEER FOUND A BURNT TRACE ON THE MAINS POWER SUPPLY OUTLET. ON EXAMINATION OF THE POWER CORD THE FIELD SERVICE ENGINEER FOUND THAT THE PIN OF THE PLUG WAS PHYSICALLY BENT. PHOTOGRAPHS OF THE DAMAGED COMPONENT HAVE BEEN SUBMITTED FOR EXAMINATION TO KEYMED. KEYMED AS THE MFR (OF THE WORKSTATION) HAVE REQUESTED THE RETURN OF THE POWER CABLE AND PLUG FOR A FULL INVESTIGATION. HOWEVER, THE DAMAGE SHOWN IN PHOTO IS CONSISTENT WITH A BROKEN PIN INSIDE THE PLUG MOULDING, MOST PROBABLY CAUSED BY ATTEMPTING TO MOVE THE WORKSTATION TO ANOTHER LOCATION WHILE IT IS STILL CONNECTED TO THE MAINS SUPPLY. THIS EVENT IS ALREADY CAUTIONED AGAINST IN THE INSTRUCTIONS FOR USE FOR THE WORKSTATION. THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
THE OLYMPUS MOBILE WORKSTATION IS DESIGNED TO BE USED WITH A RANGE OF OLYMPUS EQUIPMENT TO FACILITATE GI ENDOSCOPY, ENDOSCOPIC ULTRASOUND, RESPIRATORY AND SURGICAL ENDOSCOPIC PROCEDURES. THERE WAS NO PT INVOLVEMENT IN THIS EVENT. THERE WAS NO PROCEDURE TAKING PLACE DURING THIS EVENT THIS EVENT TOOK PLACE DURING AN INSPECTION BY AN OLYMPUS FIELD SERVICE ENGINEER AT THE HEALTHCARE FACILITY. THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION. THE FOLLOWING DEVICES WERE ON THE WORKSTATION: CV-260 VIDEO SYSTEM CENTER, CLV-206 LIGHT SOURCE, AN OEP-4 VIDEO PRINTER, IT-1, OEV191H LCD DISPLAY, AND A SONY DVD. WHEN THE FIELD SERVICE ENGINEER TURNED ON THE POWER TO THE LIGHT SOURCE, SMOKE OCCURRED BETWEEN THE PLUS OF THE WORKSTATION TRANSFORMER AND THE MAIN POWER SUPPLY SOCKET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 579669 | OLYMPUS KEYMED | OLYMPUS KEYMED MOBILE WORKSTATION | FEM | KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT), LTD. | WM-NP1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | AN OEP-4 VIDEO PRINTER| CLV-206 LIGHT SOURCE| OEV191H LCD DISPLAY| CV-260 VIDEO SYSTEM CENTER| SONY DVD| IT-1 |