FDA Adverse Event Malfunction Summary report: N

ADVIA CENTAUR XP FOLATE (FOL) ASSAY

MDR report key: 3566494 · Received January 9, 2014

Report

Report Number
1219913-2014-00006
Event Type
Malfunction
Date Received
January 9, 2014
Date of Event
December 11, 2013
Report Date
December 12, 2012
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
CGN
PMA / PMN Number
K010050
Removal / Correction Number
1219913-03/25/2014-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SIEMENS FILED THE INITIAL MDR 1219913-2014-00006 ON JANUARY 9, 2014 FOR AN INCREASE IN PATIENT RUNNING MEAN FOR ADVIA CENTAUR XP FOLATE WHOLE BLOOD CELL SAMPLES WHEN USING REAGENT LOT 071223 WHEN COMPARED TO PREVIOUSLY USED REAGENT LOTS. APRIL 9, 2014 - ADDITIONAL INFORMATION CUSTOMER PATIENT RUNNING MEAN PER REAGENT LOT. RESULTS DO NOT INCLUDE THE WHOLE BLOOD CELL CALCULATION. REAGENT LOT NUMBER: 071211, NG/ML: 8.33 , NMOL/L: 18.87; 071212, 8.84, 20.02; 071213, 7.5, 16.99; 071214, 7.49, 16.96; 071215, 8.9, 20.16; 071216, 9.06, 20.52; 071217, 8.38, 18.98; 071218, 7.4, 16.76; 071219, 9.28, 21.02; 071222, 12.54, 28.40; 071223, 12.69, 28.74; 071225, 12.26 , 27.77. BASED UPON ADDITIONAL INFORMATION AND INVESTIGATION, ON MARCH 25, 2014, SIEMENS ISSUED AN URGENT MEDICAL DEVICE RECALL (108180053) TO ITS CUSTOMERS IN THE UNITED STATES AND AN URGENT FIELD SAFETY NOTICE (10818055) TO ITS CUSTOMERS OUTSIDE OF THE UNITED STATES, RELATED TO THE ADVIA CENTAUR FOLATE ASSAY. SIEMENS HAS DETERMINED THAT THE ADVIA CENTAUR SYSTEMS FOLATE ASSAY, KIT LOTS ENDING IN 222, 224, 225, 226 AND/OR 227, MEETS THE PUBLISHED CLAIMS IN THE IFU FOR NORMAL WHOLE BLOOD CELL FOLATE. THE WHOLE BLOOD CELL FOLATE RESULTS FOR LOTS ENDING IN 219 MAY SHOW A NEGATIVE BIAS OF UP TO (B)(4) RELATIVE TO NEW LOTS 224, 225, 226, AND 227. THIS NEGATIVE BIAS MAY RESULT IN A PATIENT RECEIVING FOLATE SUPPLEMENTATION; HOWEVER, THIS SUPPLEMENTATION DOES NOT CONSTITUTE A RISK TO HEALTH.

Additional Manufacturer Narrative · 1

THE CAUSE FOR THE OBSERVED INCREASE IN THE PATIENT RUNNING MEAN FOR ADVIA CENTAUR XP FOLATE WHEN USING REAGENT LOT 071223 WHEN COMPARED TO PREVIOUS REAGENT LOTS IS UNKNOWN AND IS CURRENTLY UNDER INVESTIGATION. ADDITIONAL REPRESENTATIVE DATA HAS BEEN REQUESTED BY SIEMENS. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER OBSERVED AN INCREASE IN THE PATIENT RUNNING MEAN FOR ADVIA CENTAUR XP FOLATE WHEN USING REAGENT LOT 071223 WHEN COMPARED TO PREVIOUS REAGENT LOTS. THERE ARE NO KNOWN REPORTS OF PATIENT TREATMENT BEING ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE INCREASED ADVIA CENTAUR XP FOLATE RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17068 ADVIA CENTAUR XP FOLATE (FOL) ASSAY FOLATE IMMUNOASSAY CGN SIEMENS HEALTHCARE DIAGNOSTICS, INC. N/A 071223

Patients

Seq Age Sex Outcome Treatment
1