FDA Adverse Event Malfunction Summary report: N

MCKESSON ANESTHESIA CARE

MDR report key: 3565659 · Received October 10, 2013

Report

Report Number
3009662297-2013-00002
Event Type
Malfunction
Date Received
October 10, 2013
Date of Event
August 12, 2013
Report Date
October 10, 2013
Manufacturer
MCKESSON TECHNOLOGIES INC.
Product Code
BSZ
PMA / PMN Number
K113633
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON INVESTIGATION, IT WAS DETERMINED THAT IF THE USER'S PASSWORD CONTAINS ANY OF THE FOLLOWING CHARACTERS: PLUS SIGN (+), SINGLE RIGHT QUOTE ('), BACKSPACE (\), DATA ENTERED OR EDITED BY THE INDIVIDUAL USER ON THE MEDICAL HISTORY SCREEN OF THE PRE-ANESTHESIA EVALUATION (PAE) MODULE IS NOT RETAINED. THIS INFO WAS REPORTED BACK TO THE REPORTING ANESTHESIOLOGIST; HE CONFIRMED THE USE OF THE CHARACTERS IN HIS PASSWORD AND CHANGED HIS PASSWORD. BASED ON THE UPDATE OF THE USER'S PASSWORD, THE MCKESSON ANESTHESIA CARE (MAC) SOFTWARE APPLICATION FUNCTIONED AS INTENDED FOR THE USER. A HISTORICAL REVIEW OF COMPLAINTS INDICATED THAT THIS IS THE FIRST REPORTED COMPLAINT RECEIVED OF THIS NATURE. THERE WAS NO INJURY ON THE ACTUAL PT AND/OR NO DELAY IN PT CARE.

Description of Event or Problem · 1

IT WAS REPORTED BY ONE CUSTOMER SITE OF AN OCCURRENCE WHERE A SINGLE USER WAS UNABLE TO RETAIN ANY MEDICAL HISTORY COMMENTS IN THE MEDICAL HISTORY SCREEN OF PRE-ANESTHESIA EVALUATION (PAE) APPLICATION. THERE WERE NO ADVERSE EVENTS OR PT INJURIES RESULTANT FROM THE REPORTED ALLEGED MALFUNCTION.