NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2014-00003
- Event Type
- Malfunction
- Date Received
- January 6, 2014
- Report Date
- December 16, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO OUR FISHER & PAYKEL HEALTHCARE (FPH) SERVICE CENTRE IN THE US, WHERE IT WAS VISUALLY INSPECTED AND PERFORMANCE TESTED BY A TRAINED FPH SERVICE ENGINEER. OUR ANALYSIS IS BASED ON THE EVENT DESCRIPTION AND SERVICE REPORT PROVIDED BY THE FPH SERVICE ENGINEER. RESULTS: THE HEALTHCARE FACILITY REPORTED THAT THE PRESSURE ON THE SUBJECT NEOPUFF UNIT WAS NOT REACHING THE PROPER PRESSURE. UPON INSPECTION, IT WAS FOUND THAT THE VALVE SYSTEM WAS FAULTY. IT WAS ALSO OBSERVED THAT THE UPPER END CAP, THE LOWER END CAP AND THE FASCIA WERE BROKEN. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090721. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. IT IS MOST LIKELY THAT THE DAMAGE OBSERVED TO THE NEOPUFF VALVE WAS DUE TO IMPACT. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT." THE NEOPUFF UNIT WAS RETURNED TO THE HEALTHCARE FACILITY ONCE THE VALVE SYSTEM HAD BEEN REPLACED AND HAD PASSED THE PERFORMANCE TESTS SPECIFIED IN OUR NEOPUFF TECHNICAL MANUAL.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT AN RD900AFU NEOPUFF INFANT RESUSCITATOR WOULD NOT REACH THE CORRECT PRESSURE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 6317 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LIMITED | RD900 | 090721 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |