FDA Adverse Event Death Summary report: N

BREATHING BAG

MDR report key: 354674 · Received September 28, 2001

Report

Report Number
1824226-2001-00001
Event Type
Death
Date Received
September 28, 2001
Report Date
September 28, 2001
Manufacturer
KING SYSTEMS CORP.
Product Code
BTC
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT HAS BEEN VERBALLY REPORTED TO KING SYSTEMS OF A PT DEATH AT HOSP, WHERE KING SYSTEMS BREATHING CIRCUITS AND BREATHING BAGS ARE USED. THE INCIDENT, INVOLVING A PT, WAS DESCRIBED AS A TRAUMA CASE. THE PT WAS REPORTEDLY ON "HIGH VENTILATORY PRESSURES". STAFF INVOLVED WITH THE CASE INDICATED THAT THEY HEARD A "POP". THEY CHECKED THE BREATHING BAG, AND REPORTED THAT THERE WAS A "HOLE IN IT". THEY THEN TRIED USING TWO ADD'L BREATHING BAGS ON THE EQUIPMENT WITH THE SAME RESULTS. THE BREATHING BAGS ALLEGEDLY INVOLVED IN THE INCIDENT WERE NOT RETAINED. HOWEVER, A SAMPLE WITH THE SAME LOT NUMBER AND PRODUCT CODE WAS SENT FROM THE FACILITY TO KING SYSTEMS FOR TESTING. TEST RESULTS INDICATED NO DEFECTS. THE FACILITY SPOKESPERSON DID NOT ATTRIBUTE THE DEATH TO THE KING SYSTEMS BREATHING BAG(S). NO ADD'L PT OR PRODUCT INFO HAS BEEN SUPPLIED TO KING SYSTEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
43954 BREATHING BAG RESERVIOR BAG BTC KING SYSTEMS CORP. NA 13WU9

Patients

Seq Age Sex Outcome Treatment
1 * Death