FDA Adverse Event Injury Summary report: N

BALLOON PUMP

MDR report key: 35334 · Received July 12, 1996

Report

Report Number
35334
Event Type
Injury
Date Received
July 12, 1996
Date of Event
March 7, 1996
Report Date
March 15, 1996
Manufacturer
DATASCOPE CORP.
Product Code
DSP
Adverse Event
Yes
Report Source
User Facility report
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

AUTO FILL FAILURE, PUMP QUIT ON PT. CALLED CO TO CHECK. CO FOUND BROKEN IMPELLOR ON KNF MOTOR ASSY. CAUSED MOTOR TO OVERHEAT. CO'S SVC FOUND MOTOR FAN INTERNAL BROKEN OFF SHAFT CAUSING MOTOR TO OVERHEAT AND SHUT DOWN. TESTED UNIT OK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BALLOON PUMP BALLOON PUMP DSP DATASCOPE CORP. SYSTEM 90 NA

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Life Threatening