FDA Adverse Event
Injury
Summary report: N
BALLOON PUMP
MDR report key: 35334
·
Received July 12, 1996
Report
- Report Number
- 35334
- Event Type
- Injury
- Date Received
- July 12, 1996
- Date of Event
- March 7, 1996
- Report Date
- March 15, 1996
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
AUTO FILL FAILURE, PUMP QUIT ON PT. CALLED CO TO CHECK. CO FOUND BROKEN IMPELLOR ON KNF MOTOR ASSY. CAUSED MOTOR TO OVERHEAT. CO'S SVC FOUND MOTOR FAN INTERNAL BROKEN OFF SHAFT CAUSING MOTOR TO OVERHEAT AND SHUT DOWN. TESTED UNIT OK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BALLOON PUMP | BALLOON PUMP | DSP | DATASCOPE CORP. | SYSTEM 90 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Life Threatening |