FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 3533322 · Received December 23, 2013

Report

Report Number
9611451-2013-01075
Event Type
Malfunction
Date Received
December 23, 2013
Date of Event
November 25, 2013
Report Date
November 26, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. THE INSPIRATORY AND EXPIRATORY HEATER WIRES WERE RESISTANCE TESTED USING A CALIBRATED MULTIMETER. RESULTS: THE RESISTANCE TEST REVEALED THAT THE INSPIRATORY HEATER WIRE WAS OUT OF SPECIFICATION. NO FAULT WAS FOUND WITH THE EXPIRATORY HEATER WIRE. CONTINUITY TESTING AND VISUAL INSPECTION REVEALED THAT THE OPEN CIRCUIT IN THE INSPIRATORY HEATER WIRE WAS LOCATED NEAR THE HEATER WIRE RETAINER. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 130814. CONCLUSION: WE ARE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE OBSERVED DAMAGE ON THE INSPIRATORY HEATER WIRE. RESISTANCE TESTS AND VISUAL INSPECTIONS ARE PERFORMED ON ALL BREATHING CIRCUITS DURING PRODUCTION. ANY CIRCUIT THAT FAILS IS REJECTED. THIS SUGGESTS THAT THE HEATER WIRE BECAME OPEN CIRCUIT AFTER IT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS". OUR MONITORING AND TRENDING OF COMPLAINTS INVOLVING ELECTRICAL OPEN CIRCUITS IN HEATER WIRES IN ADULT BREATHING CIRCUITS HAS A RATE OF OCCURRENCE OF 20 DEVICES PER MILLION SOLD WORLDWIDE IN THE LAST YEAR TO THE END OF (B)(6) 2013.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT AN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS CAUSING THE MR850 TO ALARM WHEN THE HEATER WIRE ADAPTOR WAS CONNECTED. THIS WAS FOUND DURING THE SETUP CHECK AND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673229 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 130814

Patients

Seq Age Sex Outcome Treatment
1 FPH MR850 RESPIRATORY HUMIDIFIER