FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3533175 · Received December 23, 2013

Report

Report Number
2531779-2013-24972
Event Type
Malfunction
Date Received
December 23, 2013
Report Date
December 12, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/12/2013 WITH THE FOLLOWING FINDINGS:DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED. ADDITIONALLY, THE KEYPAD COVER WAS TORN FROM THE UP ARROW BUTTON TO THE CONTRAST BUTTON, AND THE CONTRAST BUTTON WAS UNRESPONSIVE, WHICH HAS NO EFFECT ON INSULIN DELIVERY. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER THE CONTRAST, UP ARROW, AND DOWN ARROW KEY CONTACTS. (B)(4).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISPLAY FAILURE AND CONTAMINATION UNDER THE KEYPAD BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 12/12/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673723 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 36 YR