FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3533048 · Received December 23, 2013

Report

Report Number
2531779-2013-24957
Event Type
Malfunction
Date Received
December 23, 2013
Report Date
December 9, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/09/2013 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE TOP SLOT OF THE BATTERY CAP WAS DAMAGED. THE BATTERY CAP THREADS WERE ALSO STRIPES, CAUSING THE BATTERY CAP TO GET STUCK ON THE PUMP. A TEST BATTERY CAP WAS USING FOR INVESTIGATION, AND THE PUMP FUNCTIONED NORMALLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.(B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE BATTERY CAP WAS DAMAGED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 12/09/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673487 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR