FDA Adverse Event
Malfunction
Summary report: N
BIRD 6400 ST VENTILATOR
MDR report key: 35323
·
Received July 23, 1996
Report
- Report Number
- 35323
- Event Type
- Malfunction
- Date Received
- July 23, 1996
- Date of Event
- July 11, 1996
- Report Date
- July 23, 1996
- Manufacturer
- BIRD PRODUCTS CORP.
- Product Code
- CBK
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
PT WAS BEING VENTILATED IN SIMV MODE. PT BEING SUCTIONED WHEN MACHINE SHUT DOWN; PT UNHARMED. MED STAFF PRESENT AND PROVIDED MANUAL RESPIRATION ALONG WITH SPONTANEOUS RESP. SVC REP REPORTED THAT INSULATION FROM CABINET HAD BECOME DISLODGED AND BLOCKED. COOLING FAN CAUSING OVERHEATING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIRD 6400 ST VENTILATOR | VENTILATOR | CBK | BIRD PRODUCTS CORP. | 09611 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |