VENTED AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2013-01053
- Event Type
- Malfunction
- Date Received
- December 22, 2013
- Date of Event
- November 27, 2013
- Report Date
- December 2, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(6) FOR EVALUATION. IT WAS VISUALLY INSPECTED AND CONNECTED TO A WATER BAG AND PERFORMANCE TESTED. RESULTS: UPON CONNECTING TO A WATER BAG A DROP OF WATER BEGAN TO BUILD AT THE CONNECTION BETWEEN THE FEEDSET TUBE AND THE BAG SPIKE CONNECTION ON THE RETURNED CHAMBER. VISUAL INSPECTION IDENTIFIED THAT SUFFICIENT GLUE WAS PRESENT HOWEVER THE GLUE DID NOT BOND PROPERLY. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT DATE 130614. CONCLUSION: WE ARE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED FAULT. WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE AT THE CHAMBER DOME IN ANY OF OUR TESTING. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. IF ANY CHAMBER FAILS, THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. ADDITIONALLY, ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THIS SUGGESTS THAT THE PROBLEM OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." OUR MONITORING AND TRENDING OF COMPLAINTS OF LOOSE WATER BAG SPIKES ON MR290 FEEDSETS HAS A RATE OF OCCURRENCE OF (B)(4) DEVICES PER (B)(4) SOLD WORLDWIDE IN THE LAST YEAR TO THE END OF NOVEMBER 2013.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT WATER WAS LEAKING FROM THE CONNECTION BETWEEN THE BAG SPIKE AND THE WATER FEEDSET TUBE ON AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND AFTER ONE DAY OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 671111 | VENTED AUTOFEED HUMIDIFICATION CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 130614 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |