DIMENSION(R) CLINICAL CHEMISTRY SYSTEM -
Report
- Report Number
- 2517506-2013-00425
- Event Type
- Malfunction
- Date Received
- December 20, 2013
- Date of Event
- February 19, 2013
- Report Date
- November 25, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- MKW
- PMA / PMN Number
- K023065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
THE CAUSE OF THE QC VARIABILITY AND LOW LEVEL QC OUT OF RANGE RECOVERY IS LIKELY RELATED TO POOR OPEN WELL STABILITY OF THE CSA REAGENT. A SHARED CHROME PARTICLE REAGENT IN COMMON WITH THE DIMENSION VISTA CSA REAGENT, (B)(4), IS THE LIKELY CAUSE. SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED COMPLAINTS REGARDING VARIABILITY OF RECOVERY OF QC AND PATIENT RESULTS WITH THE (B)(4) FLEX REAGENT CARTRIDGE LOTS 12300BB, 12318BB, AND 13011BB. INTERNAL TESTING HAS CONFIRMED ATYPICAL PERFORMANCE AND INCREASES OF >25% OVER SEVERAL DAYS FOR SAMPLES WITH CSA CONCENTRATIONS BELOW 125 NG/ML. DAILY QC CAN DETECT PERFORMANCE VARIATIONS IN THE ASSAY. ERRORS ARE MOST PRONOUNCED BELOW 125 NG/ML WHERE MAINTENANCE THERAPY OCCURS, THEREFORE, DOSAGE ADJUSTMENT BASED ON A SINGLE RESULT IS EXTREMELY UNLIKELY. NEVERTHELESS, SIEMENS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION 13-90 TO CUSTOMERS WHO HAD ORDERED THESE LOTS ADVISING THEM TO DISCONTINUE USE AND TO DISCARD THE DIMENSION VISTA LOTS.
CUSTOMER REPORTED VARIABILITY OF RECOVERY OF QC CYCLOSPORIN A (CSA). THE CUSTOMER REPORTED LOWER RESULTS THAN THE THEORETICAL CONTROL MEAN FOR THEIR LOW LEVEL CONTROL PRODUCT. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED DUE TO THE VARIABLE LOW LEVEL CONTROL RECOVERY. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT THE VARIABILITY OF RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670405 | DIMENSION(R) CLINICAL CHEMISTRY SYSTEM - | CYCLOSPORINE FLEX® REAGENT CARTRIDGE | MKW | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | FA3312 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |