CARESTREAM VUE PACS
Report
- Report Number
- 1317307-2013-00018
- Event Type
- Other
- Date Received
- November 14, 2013
- Date of Event
- October 16, 2013
- Report Date
- November 13, 2013
- Manufacturer
- CARESTREAM HEALTH, INC.
- Product Code
- LLZ
- PMA / PMN Number
- K083673
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER IDENTIFIED SPECIFIC STUDIES THAT COULD NOT BE ACCESSED. CARESTREAM HEALTH, INC BEGAN AN INVESTIGATION OF THE ISSUE. THE INVESTIGATION INCLUDED REVIEW OF LOGS AND REMOTE EVAL OF THE SYSTEM. CARESTREAM'S INVESTIGATION OF THE DEVICE AND SERVER LOGS REVEALED THREE FACTORS THAT MAY HAVE CONTRIBUTED TO THE LOSS OF STUDIES AND STUDY IMAGES. PART OF THE INVESTIGATION WAS TO DETERMINE THE NUMBER OF AFFECTED STUDIES/IMAGES. AFFECTED STUDIES WERE STORED IN 2012, DURING AN INSTALLATION/UPGRADE OF THE PRIMARY PRODUCTION SERVER. DURING THE UPGRADE PERIOD THE BACKUP SERVER WAS ACTING AS THE PRODUCTION SYSTEM FOR THE SITE, RECEIVING STUDIES DIRECTLY FROM THE MODALITIES. THIS ACTING PRODUCTION SERVER WAS CONFIGURED TO COPY STUDIES TO THE PRIMARY SERVER, WHICH WOULD THEN ARCHIVE THE STUDIES. THE BACKUP SERVER WAS USING A "COPY RULE" ALONE IN THE INFO ROUTER SOFTWARE TO MOVE STUDIES TO THE PRIMARY SERVER WHERE THEY WERE THEN TO BE ARCHIVED IN LONG TERM STORAGE. WITHOUT AN ADDED "BACKUP RULE" THERE WAS NOT A 'STUDY_NEED_BACKUP' INDICATOR. THE WORKFLOW AT THE SITE WAS SET TO ALLOW DELETION OF UNARCHIVED IMAGES. WITH THAT WORKFLOW DISABLED, THE DELETION CRITERIA WOULD RESULT IN THE 'STUDY_NEED_BACKUP' INDICATOR. THE AFFECTED STUDIES/IMAGES WERE PRIOR EXAMS. WHILE THE STUDY IMAGES WERE NO LONGER AVAILABLE, THE REPORTS AND DIAGNOSES FOR THESE STUDIES WERE STILL AVAILABLE THROUGH THE PACS CLIENT.
ON (B)(6) 2013, CARESTREAM HEALTH RECEIVED A REPORT OF PRIOR IMAGES THAT COULD NOT BE ACCESSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 592154 | CARESTREAM VUE PACS | PACS | LLZ | CARESTREAM HEALTH, INC. | VERSION 11.3 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARESTREAM 10 LNG TRM SORAGE |