- Manufacturer
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Product Code
- KPS
- PMA / PMN Number
- K093514
- Removal / Correction Number
- 9613299-06/20-2013-004-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS ISSUE WAS IDENTIFIED AS PRESENTING A DIFFERENT FAILURE MODE THAN THE ONE IDENTIFIED IN MDR# 9613299-2013-00001. BASED ON ENGINEERING ANALYSIS THESE LOOSE SCREWS DO NOT COMPROMISE THE STRUCTURAL INTEGRITY OF THE COMPONENT AND THEREFORE DO NOT INTRODUCE A HAZARDOUS SITUATION. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
Description of Event or Problem · 1
DURING SERVICE ACTIVITY RELATED TO A FIELD ACTION, A GE HEALTHCARE FIELD ENGINEER FOUND LOOSE SCREWS ON THE LINEAR RAILS. NO ASSOCIATED GAPS WERE MEASURED BETWEEN THE RAILS AND ROTOR. NO DETECTOR FALL EVENT AND NO INJURY OCCURRED.