FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA(R) CLINICAL CHEMISTRY SYSTEM

MDR report key: 3528819 · Received December 19, 2013

Report

Report Number
2517506-2013-00427
Event Type
Malfunction
Date Received
December 19, 2013
Date of Event
February 21, 2013
Report Date
November 25, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
MKW
PMA / PMN Number
K062236
Removal / Correction Number
2517506-12-05-2013-010-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED COMPLAINTS REGARDING VARIABILITY OF RECOVERY OF QC AND PATIENT RESULTS WITH THE CSA K4089 FLEX REAGENT CARTRIDGE LOTS 12300BB, 12318BB, AND 13011BB. INTERNAL TESTING HAS CONFIRMED ATYPICAL PERFORMANCE AND INCREASES OF >25% OVER SEVERAL DAYS FOR SAMPLES WITH CSA CONCENTRATIONS BELOW 125 NG/ML. DAILY QC CAN DETECT PERFORMANCE VARIATIONS IN THE ASSAY. ERRORS ARE MOST PRONOUNCED BELOW 125 NG/ML WHERE MAINTENANCE THERAPY OCCURS, THEREFORE, DOSAGE ADJUSTMENT BASED ON A SINGLE RESULT IS EXTREMELY UNLIKELY. NEVERTHELESS, SIEMENS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION 13-90 TO CUSTOMERS WHO HAD ORDERED THESE LOTS ADVISING THEM TO DISCONTINUE USE AND TO DISCARD THE LOTS. SIEMENS OFFERED ALTERNATE NON-IMPACTED LOTS FOR CUSTOMER REPLACEMENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED VARIABILITY OF RECOVERY OF QC CYCLOSPORIN A (CSA) DEPENDING UPON LENGTH OF TIME RUNNING THE SAMPLES WITHIN THE REAGENT OPEN WELL CLAIM. THE CUSTOMER REPORTED LOWER RESULTS WITH A LOT WHICH HAD BEEN HYDRATED DAYS BEFORE. THERE WERE NO REPORTS OF PATIENT TREATMENT ALTERED OR PRESCRIBED DUE TO THE VARIABLE RESULTS WITH TIME AFTER HYDRATION OF THE CSA REAGENT OPEN WELLS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT THE VARIABILITY OF RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666755 DIMENSION VISTA(R) CLINICAL CHEMISTRY SYSTEM CYCLOSPORINE FLEX® REAGENT CARTRIDGE MKW SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW 12318BB

Patients

Seq Age Sex Outcome Treatment
1