FDA Adverse Event Death Summary report: N

AVANTA FLUID MANAGEMENT FLUID SYSTEM

MDR report key: 3528009 · Received December 17, 2013

Report

Report Number
2520313-2013-00070
Event Type
Death
Date Received
December 17, 2013
Date of Event
October 25, 2013
Report Date
November 25, 2013
Manufacturer
MEDRAD
Product Code
DXT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

A BAYER R & I SERVICE PERFORMED AN INJECTOR CHECKOUT OF THE AVANTA FLUID MANAGEMENT SYSTEM AND CONFIRMED THE INJECTOR PERFORMED TO SPEC. THE DISPOSABLE FROM THE ALLEGED INCIDENT WERE DISCARDED AND NOT AVAILABLE FOR EVAL. THE SITE PROVIDED THE LOT NUMBER OF THE SINGLE-PT DISPOSABLE SET (SPAT) THAT WAS IN USE DURING THE ALLEGED INCIDENT. BAYER R & I QUALITY ASSURANCE PRODUCT ANALYSIS EXAMINED A RETAINED SPAT SAMPLE FROM LOT NUMBER 132201. NO VISUAL OR FUNCTIONAL DEFECTS WERE IDENTIFIED - THE PRODUCT PERFORMED TO SPEC. THE AVANTA FLUID MANAGEMENT SYSTEM HAS BEEN IN USE DAILY AT THE SITE SINCE THE REPORTED OCCURRENCE. ADD'L APPLICATION TRAINING WAS PERFORMED ON (B)(4) 2013.

Description of Event or Problem · 1

THE SITE REPORTED THE FOLLOWING TO A BAYER R & I REP: (B)(6) MALE PT WITH A HISTORY OF A MYOCARDIAL INFARCTION UNDERWENT A PERCUTANEOUS CORONARY INTERVENTION (PCI) OF THE RIGHT CORONARY ARTERY (RCA) AND SUFFERED AN ALLEGED AIR INJECTION WHILE CONNECTED TO THE AVANTA FLUID MANAGEMENT SYSTEM. THE PT WAS REPORTED TO BE STABLE AFTER THE ALLEGED AIR INJECTION. DURING STENT PLACEMENT, THE STENT DISLOCATED FROM THE BALLOON WITHOUT EXPANDING. THE SITE ATTEMPTED TO RETRIEVE THE STENT, HOWEVER, DURING THEIR RETRIEVAL EFFORTS THE AORTA DISSECTED AND THE PT BECAME UNSTABLE. THE PT WAS TREATED WITH CARDIOPULMONARY RESUSCITATION (CPR) AND EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO). THE PT WAS IMMEDIATELY SENT TO SURGERY AND DIED 2 DAYS FOLLOWING THE OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
658812 AVANTA FLUID MANAGEMENT FLUID SYSTEM ANGIOGRAPHIC INJECTOR DXT MEDRAD AVA 500 TABL

Patients

Seq Age Sex Outcome Treatment
1 60 YR Death