FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 3526948 · Received December 18, 2013

Report

Report Number
9611451-2013-01052
Event Type
Malfunction
Date Received
December 18, 2013
Date of Event
November 26, 2013
Report Date
November 27, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE TWO COMPLAINT RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT WERE RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4). ONE CIRCUIT WAS FROM LOT 130701 (MANUFACTURED 1 JULY 2013) AND THE OTHER CIRCUIT WAS FROM LOT 130730 (MANUFACTURED 30 JULY 2013). BOTH CIRCUITS WERE VISUALLY INSPECTED, PRESSURE TESTED AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST RESULT REVEALED THAT BOTH CIRCUITS WERE OUT OF SPECIFICATION DUE TO LEAKAGE COMING FROM THE TWO PATIENT END CONNECTORS. THIS WAS CONFIRMED WHEN THE SUBJECT BREATHING CIRCUITS WERE IMMERSED IN THE WATER BATH. VISUAL INSPECTION REVEALED THAT THERE WAS NOT ENOUGH GLUE PRESENT IN THE PROXIMAL CONNECTORS, CAUSING THE REPORTED LEAK. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR EITHER LOT NUMBER. CONCLUSION: THE OBSERVED LEAK WAS CAUSED BY INSUFFICIENT GLUE BEING INJECTED INTO THE PROXIMAL CONNECTOR SUCH THAT THEY DID NOT FORM A PERMANENT SEAL AFTER RELEASE FOR DISTRIBUTION. ALL RT340 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE RELEASING FOR DISTRIBUTION. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. IN ADDITION, TUBE WEIGHING AND BOND STRENGTH TESTING ARE PERFORMED EVERY (B)(4). IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS. HOSPITAL STAFF CORRECTLY CHECKED THE RT340 BREATHING CIRCUITS BEFORE PATIENT USE, WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT TWO RT340 ADULT BREATHING CIRCUITS FAILED THE VELA VENTILATOR LEAK TEST THIS WAS FOUND PRIOR TO USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663566 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 130701

Patients

Seq Age Sex Outcome Treatment
1 VELA VENTILATOR