PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-05791
- Event Type
- Malfunction
- Date Received
- December 18, 2013
- Date of Event
- November 12, 2013
- Report Date
- November 20, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT THE INSTRUMENT TIPS WERE BENT. ONE GRIP IS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE IS A .035 OFFSET AT THE TIPS. THE BENT GRIP DOES NOT EXHIBIT SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT. THIS INDICATES THAT THE LIKELY CAUSE OF BENDING IS OVERLOADING AT THE TIP. THE ELECTRICAL CONTINUITY TEST PASSED. IT WAS CONCLUDED THAT THE DAMAGES TO THE INSTRUMENT'S GRIP TIP WERE LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS A FRAYED CABLE. THE PITCH UP CABLE IS FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STICK OUT AT THE WRIST. CONDUCTOR WIRES ARE INTACT AND UNDAMAGED. NO OTHER CABLE DAMAGE WAS FOUND. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS MAIN TUBE DAMAGE. THE DISTAL END OF THE MAIN TUBE EXHIBITED VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .023 - .112 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. FAILURE ANALYSIS INVESTIGATION CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S CABLE DAMAGE AND/OR MAIN TUBE DAMAGE FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI HYSTERECTOMY PROCEDURE, THE JAWS OF THE PK DISSECTING FORCEPS INSTRUMENT WERE BROKEN. NO MISSING OR FALLEN PIECES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 663023 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10130710 283 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |