FDA Adverse Event
Other
Summary report: N
INFINIA
MDR report key: 3522611
·
Received November 29, 2013
Report
- Report Number
- 9613299-2013-00044
- Event Type
- Other
- Date Received
- November 29, 2013
- Date of Event
- July 10, 2013
- Report Date
- July 10, 2013
- Manufacturer
- GE MEDICAL SYSTEMS F.I. HAIFA
- Product Code
- KPS
- PMA / PMN Number
- K022960
- Removal / Correction Number
- 9613299-06/20/2013-004-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THIS ISSUE WAS IDENTIFIED AS PRESENTING A DIFFERENT FAILURE MODE THAN THE ONE IDENTIFIED IN MDR# 9613299-2013-00001. BASED ON ENGINEERING ANALYSIS THESE MEASURED GAPS DO NOT COMPROMISE THE STRUCTURAL INTEGRITY OF THE COMPONENT AND THEREFORE DO NOT INTRODUCE A HAZARDOUS SITUATION. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
Description of Event or Problem · 1
DURING SERVICE ACTIVITY RELATED TO A FIELD ACTION, A GE HEALTHCARE FIELD ENGINEER MEASURED A GAP BETWEEN THE LINEAR RAILS AND ROTOR. NO SCREWS WERE LOOSE ON THE ASSOCIATED LINER RAILS. NO DETECTOR FALL EVENT AND NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 621801 | INFINIA | SYSTEM, TOMOGRAPHY, COMPUTED, EMISSION | KPS | GE MEDICAL SYSTEMS F.I. HAIFA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |