Narratives

Additional Manufacturer Narrative · 1

THIS ISSUE WAS IDENTIFIED AS PRESENTING A DIFFERENT FAILURE MODE THAN THE ONE IDENTIFIED IN MDR# 9613299-2013-00001. BASED ON ENGINEERING ANALYSIS THIS LOOSE SCREWS ON THE LINEAR BEARINGS DO NOT COMPROMISE THE STRUCTURAL INTEGRITY OF THE COMPONENT AND THEREFORE DO NOT INTRODUCE A HAZARDOUS SITUATION. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

DURING SERVICE ACTIVITY RELATED TO A FIELD ACTION, A GE HEALTHCARE FIELD ENGINEER FOUND LOOSE SCREWS ON THE LINEAR BEARINGS. NO DETECTOR FALL EVENT AND NO INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621765 MILLENNIUM VG SYSTEM, TOMOGRAPHY, COMPUTED, EMISSION KPS GE MEDICAL SYSTEMS ISREAL

Patients

Seq Age Sex Outcome Treatment
1