ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-24236
- Event Type
- Malfunction
- Date Received
- December 13, 2013
- Report Date
- December 2, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/02/2013 WITH THE FOLLOWING FINDINGS:THE KEYPAD WAS FOUND TO BE TORN OVER THE OK BUTTON. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS WERE RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED KEYPAD, DISPLAY AND BATTERY COMPARTMENT ISSUES. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 12/02/2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 655171 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |