FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3519810 · Received December 13, 2013

Report

Report Number
2531779-2013-24236
Event Type
Malfunction
Date Received
December 13, 2013
Report Date
December 2, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/02/2013 WITH THE FOLLOWING FINDINGS:THE KEYPAD WAS FOUND TO BE TORN OVER THE OK BUTTON. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS WERE RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED KEYPAD, DISPLAY AND BATTERY COMPARTMENT ISSUES. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 12/02/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
655171 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 25 YR