FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3516461
·
Received December 12, 2013
Report
- Report Number
- 2531779-2013-23918
- Event Type
- Malfunction
- Date Received
- December 12, 2013
- Report Date
- December 2, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/02/2013 WITH THE FOLLOWING FINDINGS: DURING TESTING, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. THE KEYPAD COVER HAD BEEN RETURNED TORN. THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. ADDITIONALLY, THE THREADS OF THE BATTERY CAP WERE FOUND TO BE STRIPPED. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION UNDER KEY CONTACTS, A DISPLAY SCREEN FAILURE, AND BATTERY CAP DAMAGE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 12/02/2013. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 652199 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |