UNICEL® DXC 880I SYNCHRON ® ACCESS CLINICAL SYSTEM
Report
- Report Number
- 2050012-2013-00812
- Event Type
- Malfunction
- Date Received
- December 11, 2013
- Date of Event
- November 22, 2013
- Report Date
- November 22, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- MZV
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(6). A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE EVALUATED THE INSTRUMENT AND DISCOVERED A CRACK IN THE FLOWCELL, ADJACENT TO THE POTASSIUM (K) ELECTRODE PORT. THE FSE PROCEEDED TO REPLACE THE FLOWCELL TO RESOLVE THE ISSUE. ALTHOUGH IT IS UNKNOWN HOW THE FLOWCELL BECAME CRACKED, THE INSTRUMENT PROVIDED MULTIPLE INDICATIONS OF THE ISSUE, WITH CALIBRATION FAILURES, QUALITY CONTROL (QC) FAILURES, AND FLAGGED RESULTS. THE CUSTOMER CONTINUED TO REPORT PATIENT RESULTS DESPITE THESE INDICATORS. RESULTS: FLOWCELL.
THE CUSTOMER REPORTED OBTAINING FALSE HIGH POTASSIUM (K) RESULTS FOR FOURTEEN (14) PATIENT SAMPLES FROM THE UNICEL DXC 880I SYNCHRON ACCESS CLINICAL SYSTEM. THE CUSTOMER REPORTED THE RESULTS OUT OF THE LABORATORY AND ONE (1) PATIENT WAS TREATED BASED UPON A FALSE HIGH K RESULT. THIS REPORT REFERENCES THE OTHER THIRTEEN (13) PATIENTS INVOLVED IN THIS EVENT; PLEASE REFER TO MEDWATCH REPORT #2050012-2013-00813 FOR THE REPORT OF THE PATIENT WHO WAS TREATED. THE CUSTOMER REPEATED THE SAMPLES ON AN ALTERNATE ANALYZER AND ISSUED AMENDED REPORTS. THERE WAS NO CHANGE OR AFFECT TO TREATMENT FOR THE THIRTEEN PATIENTS REFERENCED IN THIS REPORT. QUALITY CONTROL (QC) RESULTS PRIOR TO THE EVENT EXHIBITED IMPRECISION. POTASSIUM QC PRIOR TO THE EVENT RECOVERED 8.7 MMOL/L AND REPEATED AT 7.5 MMOL/L (MEAN WAS 7.5 MMOL/L). PATIENT SAMPLES WERE THEN RUN AND K QC FOLLOWING THE SAMPLE RUN RECOVERED AT 9.3 AND 8.4 MMOL/L. THE CUSTOMER RECALIBRATED THE INSTRUMENT AND K QC RECOVERED AT 7.4 AND 7.6 MMOL/L. THE CUSTOMER THEN RAN PATIENT SAMPLES AGAIN. A REVIEW OF THE CUSTOMER'S INSTRUMENT DATA LOG INDICATED THAT THE SYSTEM HAD MULTIPLE CALIBRATION FAILURES STARTING ON (B)(6) 2013 THROUGH THE DAY OF THE EVENT. IN ADDITION TO THE ERRATIC QC AND CALIBRATION FAILURES, ON THE DAY OF THE EVENT, THE INSTRUMENT ALSO FLAGGED INTERMITTENT K RESULTS WITH A "SAMPLE REFERENCE DRIFT" FLAG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 648351 | UNICEL® DXC 880I SYNCHRON ® ACCESS CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | MZV | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |