ADULT INSPIRATORY HEATED BREATHING CIRCUITS
Report
- Report Number
- 9611451-2013-01005
- Event Type
- Malfunction
- Date Received
- December 11, 2013
- Date of Event
- November 6, 2013
- Report Date
- November 22, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT105 ADULT INSPIRATORY-HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IS SIMILAR TO A PRODUCT SOLD IN THE USA WITH A 510(K) OF K983112. METHOD: THE COMPLAINT RT105 ADULT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. THE ELECTRICAL RESISTANCE OF THE INSPIRATORY HEATERWIRE WAS TESTED USING A CALIBRATED MULTIMETER. CONTINUITY TEST AND VISUAL INSPECTION WERE ALSO CONDUCTED TO CHECK THE ELECTRICAL CONNECTION BETWEEN THE HEATERWIRE AND THE HEATERWIRE PINS. RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE RETURNED BREATHING CIRCUIT. THE ELECTRICAL RESISTANCE TEST SHOWED THE INSPIRATORY HEATERWIRE WAS OPEN CIRCUIT. CONTINUITY TEST AND VISUAL INSPECTION REVEALED THAT THE OPEN CIRCUIT IN THE INSPIRATORY HEATERWIRE WAS LOCATED AT THE CONNECTION BETWEEN THE HEATERWIRE AND THE RIGHT HEATERWIRE PIN INSIDE THE OVERMOULDED PLUG. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130626. CONCLUSION: ELECTRICAL OPEN CIRCUITS IN HEATERWIRES CAN BE ASSOCIATED WITH INSUFFICIENT CONNECTION OF THE HEATERWIRE PIN DURING PRODUCTION, SUCH THAT IT IS UNABLE TO PROVIDE CONTINUITY FOR THE FULL PERIOD OF USE. ALL RT105 BREATHING CIRCUITS ARE ELECTRICAL RESISTANCE AND CONTINUITY TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE SUBJECT INSPIRATORY HEATERWIRE BECAME OPEN CIRCUIT AFTER RELEASED FOR DISTRIBUTION, POSSIBLY AS A RESULT OF INSUFFICIENT CONNECTION. (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A "HEATERWIRE DISCONNECT" ALARM SOUNDED ON THE MR850 RESPIRATORY HUMIDIFIER WHEN AN RT105 ADULT INSPIRATORY HEATED BREATHING CIRCUIT WAS CONNECTED. THIS WAS OBSERVED AFTER ONE WEEK OF USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 648480 | ADULT INSPIRATORY HEATED BREATHING CIRCUITS | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | RT105 | 130626 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MR850 RESPIRATORY HUMIDIFIER |