FUKUDA DENSHI
Report
- Report Number
- 9611031-2013-00001
- Event Type
- Death
- Date Received
- November 18, 2013
- Date of Event
- October 16, 2013
- Report Date
- November 18, 2013
- Manufacturer
- FUKUDA DENSHI CO., LTD.
- Product Code
- MHX
- PMA / PMN Number
- K033711
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS EVENT OCCURRED SOLELY AS A RESULT OF USER ERROR AND NOT DUE TO ANY MALFUNCTION OF THE DEVICE. AND AS PER FUKUDA DENSHI CO. LTD. QA SYSTEM SOP, IS WAS DETERMINED TO BE A REPORTABLE EVENT AND THEREFORE AN FDA FORM 3500A WAS ORDERED TO BE FILED. IN ADDITION, AS A RESULT OF THIS EVENT, THE HOSPITAL STAFF WERE PROVIDED WITH ADDITIONAL INSTRUCTION AND THOROUGH EXPLANATION OF DEVICE OPERATION AND THE IMPORTANCE OF ROUTINELY CHECKING THE DEVICE POWER SOURCE AND MONITORING OF THE CENTRAL STATION.
EVENT: THIS EVENT OCCURRED IN A (B)(6) HOSPITAL ON (B)(6) 2013 AND REPORTED TO US BY THE HOSPITAL ON (B)(4) 2013. A MEDICAL STAFF NOTICED THE PT ECG WAVEFORM ON THE CENTRAL MONITOR STATION, (B)(4), SCREEN TO BE A SQUARE WAVEFORM INDICATING THE PT BEING MONITORED ON A TELEMETRY TRANSMITTER MONITOR, LX-5230, WAS OFF-LINE, OUT OF RANGE, OR NOT CONNECTED. THE MEDICAL STAFF PERSON RUSHED INTO PT ROOM AND FOUND THE PT HAD EXPIRED. UPON EXAMINATION OF THE TELEMETRY TRANSMITTER MONITOR, IT WAS DETERMINED THAT THE BATTERY POWER OF TELEMETER LX-5230 WAS EXHAUSTED. THIS EVENT WAS DUE TO USER ERROR AND NOT DUE TO ANY MALFUNCTION OF DEVICE, WHICH WAS ALSO ACKNOWLEDGED BY THE FACILITY. THE TELEMETRY TRANSMITTER MONITOR INVOLVED IN THIS EVENT IS THE FUKUDA DENSHI MODEL LX-5230 WHICH IS SOLD IN THE (B)(4) MARKET BUT WHICH IS SIMILAR TO THE FUKUDA DENSHI TELEMETRY TRANSMITTER MONITOR LX-5630 WHICH IS DISTRIBUTED AND SOLD IN THE US MARKET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 596121 | FUKUDA DENSHI | TELEMETRY TRANSMITTER MONITOR | MHX | FUKUDA DENSHI CO., LTD. | LX-5320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |