FDA Adverse Event Malfunction Summary report: N

RADIOGRAPHIC/FLUOROSCOPIC X-RAY UNIT-TABLE

MDR report key: 35154 · Received August 9, 1996

Report

Report Number
MW1009681
Event Type
Malfunction
Date Received
August 9, 1996
Date of Event
April 24, 1996
Report Date
April 29, 1996
Manufacturer
GENERAL ELECTRIC CO.
Product Code
KXJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

DURING A VENOGRAM PROCEDURE WITH THE X-RAY TABLE IN A SEMI-UPRIGHT POSITION (ESTIMATED 45 DEGREES) THE FLOATING TABLE TOP SEEMED TO BREAK LOOSE FROM THE REST OF THE TABLE ASSEMBLY AND SLIDE TOWARDS THE FLOOR. X-RAY PERSONNEL WERE ABLE TO MANUALLY CONTROL THE MOVEMENT OF THE TABLE TOP. THE TABLE WAS PLACED IN A HORIZONTAL POSITION AND THE PT SAFELY REMOVED WITHOUT INJURY. UPON INSPECTION BY EQUIPMENT SVC PERSONNEL IT WAS DETERMINED THAT A GEAR THAT HOLDS THE TABLE TOP CHAIN SLID OFF THE SHAFT OF THE MOTOR. THUS ALLOWING THE TABLE TOP TO FLOAT FREELY. THIS EQUIPMENT IS NORMALLY MAINTAINED BY A THIRD PARTY RADIOLOGY SVC VENDOR. THE THIRD PARTY VENDOR REPLACED THE GEAR BOX IN QUESTION AROUND 8/3/95.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RADIOGRAPHIC/FLUOROSCOPIC X-RAY UNIT-TABLE RADIOGRAPHIC/FLUOROSCOPIC X-RAY UNIT-TABLE KXJ GENERAL ELECTRIC CO.

Patients

Seq Age Sex Outcome Treatment
1 83 YR Other