FDA Adverse Event
Malfunction
Summary report: N
RADIOGRAPHIC/FLUOROSCOPIC X-RAY UNIT-TABLE
MDR report key: 35154
·
Received August 9, 1996
Report
- Report Number
- MW1009681
- Event Type
- Malfunction
- Date Received
- August 9, 1996
- Date of Event
- April 24, 1996
- Report Date
- April 29, 1996
- Manufacturer
- GENERAL ELECTRIC CO.
- Product Code
- KXJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
DURING A VENOGRAM PROCEDURE WITH THE X-RAY TABLE IN A SEMI-UPRIGHT POSITION (ESTIMATED 45 DEGREES) THE FLOATING TABLE TOP SEEMED TO BREAK LOOSE FROM THE REST OF THE TABLE ASSEMBLY AND SLIDE TOWARDS THE FLOOR. X-RAY PERSONNEL WERE ABLE TO MANUALLY CONTROL THE MOVEMENT OF THE TABLE TOP. THE TABLE WAS PLACED IN A HORIZONTAL POSITION AND THE PT SAFELY REMOVED WITHOUT INJURY. UPON INSPECTION BY EQUIPMENT SVC PERSONNEL IT WAS DETERMINED THAT A GEAR THAT HOLDS THE TABLE TOP CHAIN SLID OFF THE SHAFT OF THE MOTOR. THUS ALLOWING THE TABLE TOP TO FLOAT FREELY. THIS EQUIPMENT IS NORMALLY MAINTAINED BY A THIRD PARTY RADIOLOGY SVC VENDOR. THE THIRD PARTY VENDOR REPLACED THE GEAR BOX IN QUESTION AROUND 8/3/95.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RADIOGRAPHIC/FLUOROSCOPIC X-RAY UNIT-TABLE | RADIOGRAPHIC/FLUOROSCOPIC X-RAY UNIT-TABLE | KXJ | GENERAL ELECTRIC CO. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Other |