FDA Adverse Event Injury Summary report: N

OPSIS Z-CAM

MDR report key: 35041 · Received August 7, 1996

Report

Report Number
MW1009658
Event Type
Injury
Date Received
August 7, 1996
Date of Event
July 11, 1996
Report Date
July 29, 1996
Manufacturer
V. MUELLER DIV. BAXTHER HEALTHCARE CORP.
Product Code
FWC
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
GA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

SURGEONS COMPLAINED THE CAMERA WAS BLURRY AND OUT OF FOCUS, AND UNABLE TO SEE ANATOMY. LAPAROSCOPIC CHOLECYSTECTOMY CONVERTED TO OPEN PROCEDURE. BIOMEDICAL ENGINEERING TECH EVALUATION OF CAMERA SYSTEM FOUND THAT THE CAMERA PICTURE HAS TOO MUCH CHROMA (PURPLE). DURING THE PAST YEAR THE CAMERA HAS "HSD" PROBLEMS WITH COLOR CHANGING AND AN INABILITY TO MAINTAIN COLOR DURING SURGERY CASES. THE CAMERA HAS ALSO BEEN FLUCTUATING IN LIGHT INTENSITY, WHICH HAS MADE IT DIFFICULT TO DISTINGUISH THE COLOR OF TISSUE. THE HOSP OWNS 2 CAMERA SYSTEMS AND BOTH HAVE A HISTORY OF POOR QUALITY PICTURE. BOTH HAVE BEEN RETURNED TO THE MFR TWICE DURING THE PAST YEAR FOR VARIOUS PROBLEMS, THE MFR HAS NOT FOUND ANY PROBLEM WITH THE COLOR FLUCTUATIONS WITH THE CAMERAS. THE MONITORS HAVE ALSO BEEN RETURNED TO THE MFR FOR EVALUATION OF COLOR CHANGES. REPLACED A DIODE IN EACH TO IMPROVE THE PICTURE CONTRAST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPSIS Z-CAM SURGICAL CAMERA BOX FWC V. MUELLER DIV. BAXTHER HEALTHCARE CORP. VS7100 *

Patients

Seq Age Sex Outcome Treatment
1 33 YR Required Intervention