FDA Adverse Event Death Summary report: N

HEWLETT PACKARD

MDR report key: 349924 · Received August 31, 2001

Report

Report Number
349924
Event Type
Death
Date Received
August 31, 2001
Date of Event
June 14, 2001
Report Date
August 31, 2001
Manufacturer
AGILENT TECHNOLOGIES
Product Code
DRT
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

RN NOTICED THE TRACH TUBE WAS PARTIALLY DISLODGED. THE PT WAS GRAY AND UNRESPONSIVE TO STIMULI. THE TRACH TUBE WAS FOUND TO BE IN THE TRACHEOSTOMY STOMA. THE MONITOR DID NOT ALARM. THE PT EXPIRED FROM HYPOXIC ISCHEMIC ENCEPHALOPATHY. ON THE DAY OF THE EVENT THE MONITOR WAS TAKEN OUT OF SERVICE AND SENT TO THE BIOMEDICAL ENGINEERING DEPT TO BE EVALUATED. THE MONITOR WAS FOUND TO BE IN PROPER WORKING CONDITION, BUT IT HAD THE ABILITY FOR ITS ALARMS TO BE IN INFINITE SUSPEND MODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39819 HEWLETT PACKARD MONITOR DRT AGILENT TECHNOLOGIES M1275A *

Patients

Seq Age Sex Outcome Treatment
1 * Death