FDA Adverse Event Malfunction Summary report: N

SEAC DENTAL SYSTEM

MDR report key: 3495578 · Received December 2, 2013

Report

Report Number
1722638-2013-00001
Event Type
Malfunction
Date Received
December 2, 2013
Date of Event
October 1, 2013
Report Date
October 31, 2013
Manufacturer
ASI MEDICAL, INC.
Product Code
EIA
PMA / PMN Number
K974487
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED TO U.S. FOR EVAL AND TRADE IN BY THE CUSTOMER. HOWEVER, THE CUSTOMER HAD REMOVED AND DISCARDED THE PARTS OF THE AIR TANK FROM THE CART. SO WE ARE UNABLE TO EVALUATE AND DETERMINE THE SOURCE OR CAUSE OF THE FAILURE OR IF ANY EXTERNAL PHYSICAL DAMAGE MAY HAVE BEEN DONE TO IT IN ORDER TO CAUSE THE FAILURE.

Description of Event or Problem · 1

THE DENTIST HAS A MODEL 2020 MOBILE SEAC DENTAL CART WITH A BUILD IN AIR COMPRESSOR AND AIR TANK. THE CART IS ELEVEN YEARS OLD AND WAS MANUFACTURED IN OCTOBER 2002. THE AIR COMPRESSOR AND AIR TANK ARE FULLY ENCLOSED INSIDE THE METAL CHASSIS OF THE CART WHICH ARE ONLY ACCESSIBLE THROUGH A SERVICE PANEL. THE CART USED A PLASTIC AIR TANK THAT BLEW APART AFTER HOURS. THERE WERE NO INJURIES AND NO PTS INVOLVED. HOWEVER, IF A SERVICE TECHNICIAN WAS WORKING ON THE CART AND THIS PROBLEM WERE TO REOCCUR IT COULD RESULT IN BODILY INJURY TO THEIR HANDS. THIS CART MODEL DESIGN WITH THE PLASTIC AIR TANK WAS DISCONTINUED IN 2004 AND THIS POTENTIAL PROBLEM EXISTS ONLY WITH CARTS MADE ON OR BEFORE TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
622328 SEAC DENTAL SYSTEM DENTAL UNIT EIA ASI MEDICAL, INC. 2020

Patients

Seq Age Sex Outcome Treatment