NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2013-00955
- Event Type
- Malfunction
- Date Received
- November 27, 2013
- Report Date
- October 31, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO OUR FISHER & PAYKEL HEALTHCARE (FPH) SERVICE CENTER IN (B)(4), WHERE IT WAS INSPECTED BY A TRAINED FPH TECHNICIAN. RESULTS: VISUAL INSPECTION OF THE NEOPUFF REVEALED THAT THE GAS INLET PORT WAS BROKEN. THE CASING WAS ALSO CRACKED AND PRESSURE TESTING REVEALED THAT THE MANOMETER WAS FAULTY. A LOT CHECK REVEALED TWO OTHER COMPLAINTS OF THIS TYPE FOR LOT NUMBER 070208. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANT UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. IT IS MOST LIKELY THAT THE PHYSICAL DAMAGE TO THE GAS INLET PORT, CASING AND MANOMETER WAS CAUSED BY SOME SORT OF IMPACT. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF/PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT."
A HOSPITAL IN (B)(6) REPORTED THAT THE GAS INLET PORT OF A NEOPUFF INFANT RESUSCITATOR WAS BROKEN AND REQUESTED A PERFORMANCE CHECK. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 620402 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 070208 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |