BABY CONTROL COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2013-00952
- Event Type
- Malfunction
- Date Received
- November 26, 2013
- Report Date
- October 28, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE DEFECTIVE WARMER HEAD ASSEMBLY OF THE IW930JEU BABY CONTROL COSYCOT INFANT WARMER WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. OUR ANALYSIS IS ACCORDINGLY BASED ON PHOTOGRAPHS PROVIDED BY THE CUSTOMER. RESULTS: PHOTOGRAPHS SHOWED THAT THE WARMER HEAD ASSEMBLY WAS CRACKED. CRAZING AND SMALL CRACKS WERE FOUND ON THE SURFACE OF THE UPPER HEAD CASING. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 030905. CONCLUSION: IT IS LIKELY THAT THE CRACKING AND CRAZING OBSERVED ON THE WARMER HEAD ASSEMBLY ARE RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. WHEN THE WARMER HEAD BEGINS TO WARM UP DURING USE, A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRACKING AND CRAZING OF THE WARMER HEAD. THE INFANT WARMER SERVICE MANUAL FOR THE AFFECTED UNIT FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES THE FOLLOWING: "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTERALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE"; "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES."
(B)(4). WE ARE CURRENTLY ATTEMPTING TO RETRIEVE THE COMPLAINT DEVICE. WE WILL SEND A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A MEDICAL CENTRE IN (B)(6) REPORTED THAT THE UPPER HEAD CASING OF AN IW930 BABY CONTROL COSYCOT INFANT WARMER HAD CRACKED. NO PATIENT CONSEQUENCE WAS REPORTED.
A MEDICAL CENTRE IN (B)(6) REPORTED THAT THE UPPER HEAD CASING OF AN IW930 BABY CONTROL COSYCOT INFANT WARMER HAD CRACKED. NO PATIENT CONSEQUENCE WAS REPORTED.