FDA Adverse Event
Injury
Summary report: N
BIRD OXYGEN BLENDER
MDR report key: 34904
·
Received July 23, 1996
Report
- Report Number
- 34904
- Event Type
- Injury
- Date Received
- July 23, 1996
- Date of Event
- July 12, 1996
- Report Date
- July 22, 1996
- Manufacturer
- BIRD PRODUCTS CORP.
- Product Code
- BZR
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
ABG BLOOD GAS DRAWN ON 7/12/96 INDICATED ELEVATED PO2. ON INVESTIGATION OXYGEN BLENDER DIAL READ .21 WITH A FLOW OF 0.5 LPM. THE NASAL CANNULA WAS ANALYZED AND READ 40% OXYGEN. AN INVESTIGATION OF DEVICE ON 7/18/96 BY THE MFR SHOWED THIS HIGH FLOW RATE BLENDER TO BE ERRATIC IN PERFORMANCE WHEN USED TO DELIVER A LOW FLOW RATE. WEIGHT 819 GM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIRD OXYGEN BLENDER | OXYGEN BLENDER | BZR | BIRD PRODUCTS CORP. | 5100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 DAY | Required Intervention |