FDA Adverse Event Malfunction Summary report: N

THUDERBEAT 5NN, 35CM, FRONT-ACTUATED GRIP

MDR report key: 3489858 · Received September 25, 2013

Report

Report Number
8010047-2013-00490
Event Type
Malfunction
Date Received
September 25, 2013
Date of Event
July 24, 2013
Report Date
August 29, 2013
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PROBE BROKE DOWN AT 15 MM FROM THE DISTAL END. AND THERE WAS A SCRATCH AT THE UPSIDE OF THE BROKEN POINT. THE SHAPE OF THE FRACTURE SURFACE SHOWED THAT THE FRACTURE DEVELOPED FROM THE SCRATCHED AS THE STARTING POINT. THE MANUFACTURING HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. AS THE RESULT OF THE EVALUATION, WE HAVE CONCLUDED THAT THE PROBE PRESUMABLY WAS SCRATCHED BECAUSE OF THE PHYSICIAN ACTIVATED ULTRASOUND OUTPUT WHILE THE PROBE WAS IN CONTACT WITH METAL SUCH AS A CLIP AND FORCEPS OR THE PHYSICIAN SCRAPE AND PROBE WITH A SHARP OBJECT SUCH AS A SCALPEL. IN ADDITION, SINCE THE PHYSICIAN CONTINUED TO USE THE SCRATCHED DEVICE, THE CRACK OF THE PROBED OCCURRED AND ERROR DISPLAYED. CONSEQUENTLY, DURING THE EXAMINATION OUTSIDE THE PATIENT, THE PROBE TIP BROKE. THE INSTRUCTION MANUAL OF THE THUNDERBEAT SCISSORS MENTIONS WARNING AND CAUTION OF POSSIBLE MISHANDLING MENTIONED ABOVE. BASED UPON THE FINDING OF THE EVALUATION, THIS REPORT APPEARS TO BE RELATED TO USER MISHANDLING. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

THE FACILITY REPORTED THAT WHEN AN ERROR DISPLAYED DURING A PROCEDURE, THE PROBE TIP WAS BROKEN BY CLEANING AND THIS WAS OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
485516 THUDERBEAT 5NN, 35CM, FRONT-ACTUATED GRIP THUNDERBEAT HANDPIECE. GEI OLYMPUS MEDICAL SYSTEMS CORPORATION TB-0535FC K3426

Patients

Seq Age Sex Outcome Treatment
1