STERNAL CABLE Ø1 W/NEEDLE+FIXATION SLEEV
Report
- Report Number
- 8030965-2013-05345
- Event Type
- Malfunction
- Date Received
- November 26, 2013
- Date of Event
- October 25, 2013
- Report Date
- November 4, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- JDQ
- PMA / PMN Number
- K031508
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. PLACEHOLDER.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE MANUFACTURING EVALUATION REPORTED THE COMPLAINT CONDITION (UNABLE TO SEPARATE NEEDLE FROM CABLE) PERTAINS TO THE COMPONENT 291.916.1 AND IS DUE TO AN UNKNOWN CAUSE. THE 291.916.01S ASSEMBLY INITIALLY CONFORMED TO ALL REQUIREMENTS AND WAS INSPECTED AND CONFORMED TO THE SYNTHES INSPECTION REQUIREMENTS. THE MANUFACTURING EVALUATION DID NOT REPORT ANY FAULT OF THE MANUFACTURING PROCESS FOR THE COMPLAINED EVENT. PLACEHOLDER.
PRODUCT DEVELOPMENT EVALUATION WAS CONDUCTED. THE REPORT INDICATES THAT FOUR CABLES RECEIVED. IT WAS FOUND THAT THE NEEDLES COULD NOT BE REMOVED WITH FORCEPS AS PER THE DESIGN INTENT. SOME OF THE NEEDLES HAVE CLEAR MARKS FROM THE FORCEPS USED DURING THE ATTEMPT OF NEEDLE REMOVAL. NO DESIGN RELATED ROOT CAUSE CAN BE IDENTIFIED ON THE RETURNED DEVICES, THE COMPLAINT IS INDETERMINATE FROM A DESIGN PERSPECTIVE. THE DEVICE WAS RETURNED ON (B)(4) 2013.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE DEVICE HISTORY REVIEW (DHR) BY (B)(4) COMPLETED ON (B)(4) 2013 REPORTED NO COMPLAINT RELATED ISSUES WERE FOUND. IN ADDITION THE DHR REVEALED THE DEVICE WAS MANUFACTURED ON 12/8/2010 AND HAD AN EXPIRATION DATE OF 11/1/2020. THE DEVICE HISTORY REVIEW RELEVANT TO AFTER THE DEVICE WAS ASSEMBLED COMPLETED BY (B)(4) ON (B)(4) 2014 REPORTED: SYNTHES (B)(4) ASSEMBLED THE 1.0MM STERNAL CABLE WITH NEEDLE - STERILE, P/N 291.916.01S, AND LOT NUMBER 6472188) FOR WORK ORDER (B)(4), FROM LOT NUMBER 6261290 OF COMPONENT 291.916.1 AND LOT NUMBER 6221302 OF COMPONENT 291.906.1. BOTH COMPONENTS WERE MANUFACTURED BY DANIELS MANUFACTURING. LOT 6221302, OF COMPONENT 291.906.1 CONFORMED TO ALL REQUIREMENTS (C OF C DATE 9/22/09) AND WAS RELEASED FOR ASSEMBLY 9/29/09. LOT 6261290, COMPONENT 291.916.1 CONFORMED TO ALL INSPECTION REQUIREMENTS (C OF C DATE 12/18/09) AND WAS RELEASED FOR ASSEMBLY 2/3/10. THE COMPONENTS WERE ASSEMBLED AND INSPECTED TO SYNTHES INSPECTION SHEET NS025211, REVISION A (DATED 9/1/10). THERE WERE NO NON-CONFORMANCES OR OTHER COMPLAINT-RELATED ISSUES ASSOCIATED WITH THIS COMPLAINT. THE FINAL ASSEMBLY WAS RELEASED TO FINISHED GOODS 9/10/10. IN ADDITION THE DHR REVEALED THE DEVICE WAS MANUFACTURED ON 09/13/2010 AND HAD AN EXPIRATION DATE OF 08/31/2019 THE FINAL ASSEMBLY OF THE PART WAS PERFORMED AT (B)(4), THE COMPONENTS WERE STERILIZED BY (B)(4) THEREFORE TWO DHR¿S WERE PERFORMED.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2013, DURING AN UNKNOWN SURGERY THAT THE NEEDLE COULD NOT BE SEPARATED FROM THE CABLES. THE PROPER SEPARATION TECHNIQUE WAS ATTEMPTED AND WAS UNSUCCESSFUL. SUBSTANTIAL FORCE WAS USED IN ATTEMPTING TO SEPARATE THE NEEDLE FROM THE CABLES BUT WAS ALSO UNSUCCESSFUL. THE STOCK AVAILABLE INCLUDED CABLES IN WHICH THE NEEDLE COULD BE REMOVED THE NORMAL WAY, THEREFORE, THE SURGERY COULD PROCEED AS INTENDED. THIS IS REPORT 3 OF 4 FOR COMPLAINT (B)(4).
IT WAS REPORTED THAT THE CABLES USED TO COMPLETE THE SURGERY WERE FROM THE SAME LOT AS THE CABLES THAT DID NOT WORK.