VITROS DT60 II CHEMISTRY SYSTEM

MDR report key: 3485704 · Received November 25, 2013

Report

Report Number
1319681-2013-00196
Event Type
Malfunction
Date Received
November 25, 2013
Date of Event
October 28, 2013
Report Date
November 25, 2013
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT UNEXPECTED VITROS ALKP DT, CHOL DT, AND CA DT QUALITY CONTROL RESULTS WERE OBTAINED ON A VITROS DT60 II CHEMISTRY SYSTEM. THE INVESTIGATION DETERMINED THAT THE CAUSE OF THIS EVENT WAS A USER-INDUCED SLIDE TRACKING ERROR, WHICH IS A REPORTABLE MALFUNCTION FOR THE VITROS DT60 II CHEMISTRY SYSTEM. INFORMATION FROM SECTION 8.3 OF THE VITROS DT60 II OPERATOR¿S MANUAL (31-JULY-2010 REVISION) DESCRIBING THE POTENTIAL CAUSES OF A SLIDE TRACKING ERROR WAS REVIEWED WITH THE CUSTOMER. FOLLOWING ACTIONS TO RESOLVE A SLIDE TRACKING ERROR, ACCEPTABLE QUALITY CONTROL RESULTS WERE OBTAINED. THE ROOT CAUSE OF THIS EVENT IS USER ERROR.

Description of Event or Problem · 1

THE CUSTOMER OBTAINED UNEXPECTED VITROS ALKP DT, CHOL DT, AND CA DT QUALITY CONTROL RESULT WHEN USING THE VITROS DT60 II CHEMISTRY SYSTEM. WHEN RUNNING THE VITROS DT CONTROL II FLUID, A VITROS ALKP DT RESULT OF 274 U/L WAS OBTAINED VERSUS THE EXPECTED VALUE OF 401 U/L, A VITROS CHOLD DT RESULT OF 5.42 MMOL/L WAS OBTAINED VERSUS THE EXPECTED VALUE OF 6.2 MMOL/L, AND VITROS CA DT RESULTS OF >14.0 AND 2.90 MG/DL WERE OBTAINED VERSUS THE EXPECTED VALUE OF 11.7 MG/DL. NO PATIENT SAMPLES WERE RUN DURING THE TIME FRAME OF THE EVENT. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
613117 VITROS DT60 II CHEMISTRY SYSTEM CHEMISTRY ANALYZER JJE ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1