FDA Adverse Event Other Summary report: N

12 MM ENDOSCOPE, 30 DEGREE

MDR report key: 3484681 · Received November 22, 2013

Report

Report Number
2955842-2013-05390
Event Type
Other
Date Received
November 22, 2013
Date of Event
October 24, 2013
Report Date
October 24, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE 12 MM ENDOSCOPE WAS RETURNED AND EVALUATED. THE REPORTED CUSTOMER FAILURE MODE WAS CONFIRMED WITH THE EVALUATION OF THE ENDOSCOPE. THE ENDOSCOPE WAS RECEIVED WITH MECHANICAL DAMAGE TO THE DISTAL WINDOW ASSEMBLY HERMETIC SEAL RESULTING IN FLUID INVASION AND SUBSEQUENT MINOR DAMAGE TO OPTICAL COMPONENTS. INTUITIVE SURGICAL, INC. (ISI) HAS ATTEMPTED TO CONTACT THE SITE TO OBTAIN ADDITIONAL INFORMATION CONCERNING THE REPORTED EVENT; HOWEVER, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AS OF THE DATE OF THIS REPORT. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED. THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE SURGEON MADE THE DECISION TO CONVERT THE DA VINCI PROSTATECTOMY PROCEDURE TO OPEN SURGICAL TECHNIQUES AFTER ENCOUNTERING AN IMAGE ISSUE WITH THE ENDOSCOPE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE SURGICAL STAFF ENCOUNTERED AN IMAGE ISSUE WITH THE 12 MM ENDOSCOPE, 30º WHERE A HALF MOON SHAPE APPEARED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. DUE TO THE REPORTED ISSUE, THE SURGEON MADE THE DECISION TO CONVERT THE DA VINCI PROSTATECTOMY PROCEDURE TO OPEN SURGICAL TECHNIQUES. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609235 12 MM ENDOSCOPE, 30 DEGREE ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 370893-03

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES