FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3484427 · Received November 22, 2013

Report

Report Number
2955842-2013-05376
Event Type
Malfunction
Date Received
November 22, 2013
Date of Event
October 14, 2013
Report Date
October 31, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT OF DULL BLADES. THE BLADE EDGES WERE NOT DAMAGED. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES MEASURED APPROXIMATELY .041 - .135 IN LENGTH AND NOT ALIGNED WITH THE TUBE. FAILURE ANALYSIS CONCLUDED THAT THE DAMAGE MAY BE DUE TO MISHANDLING/MISUSE. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED SCRATCH MARKS FOUND DURING FAILURE ANALYSIS IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, IT WAS OBSERVED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS DULL. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
609968 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-12 M12130610 923

Patients

Seq Age Sex Outcome Treatment
1 72 YR