FDA Adverse Event Malfunction Summary report: N

PRECISE BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3484017 · Received November 22, 2013

Report

Report Number
2955842-2013-05359
Event Type
Malfunction
Date Received
November 22, 2013
Date of Event
October 1, 2013
Report Date
November 6, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. FAILURE ANALYSIS ALSO OBSERVED THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .117 - .166 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE MAY HAVE BEEN CAUSED BY MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PRECISE BIPOLAR FORCEPS INSTRUMENT WIRE WAS TORN OFF DURING A DA VINCI SURGICAL PROCEDURE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
610959 PRECISE BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420110-06 M10120731 702

Patients

Seq Age Sex Outcome Treatment
1