CATHETER MOUNT
Report
- Report Number
- 9611451-2013-00937
- Event Type
- Malfunction
- Date Received
- November 22, 2013
- Date of Event
- October 23, 2013
- Report Date
- October 23, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZO
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4).METHOD: TWO RT021 CATHETER MOUNTS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE THEY WERE VISUALLY EXAMINED. ONE OF THE RETURNED RT021 WAS FROM LOT 120926 AND THE LOT INFORMATION FOR THE OTHER RT021 WAS NOT PROVIDED.RESULTS: VISUAL INSPECTION REVEALED THAT THE TUBING CUFF AT THE CONNECTOR END OF THE CATHETER MOUNT WAS SPLIT ON THE RT021 FROM LOT 120926. THE TUBING CUFF WAS SPLIT AT THE SWIVEL END ON THE OTHER RT021 CATHETER MOUNT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120926.CONCLUSION: WE ARE UNABLE TO DETERMINE THE CAUSE OF THE DAMAGE OBSERVED ON THE RETURNED RT021 CATHETER MOUNTS. ALL RT021 CATHETER MOUNTS ARE PRESSURE TESTED PRIOR TO BEING RELEASED FOR DISTRIBUTION. ANY CATHETER MOUNT WITH A SPLIT TUBE WOULD HAVE FAILED THE PRESSURE TEST. THIS SUGGESTS THAT THE RETURNED CATHETER MOUNTS WERE DAMAGED AFTER THEY WERE RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT021 CATHETER MOUNT STATE THE FOLLOWING:"CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE";"PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT".
A DISTRIBUTOR IN (B)(6) REPORTED THAT TWO RT021 CATHETER MOUNTS WERE BROKEN. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 611072 | CATHETER MOUNT | BZO | BZO | FISHER & PAYKEL HEALTHCARE LTD | RT021 | 120926 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |