FDA Adverse Event Injury Summary report: N

EZ BREATH ATOMIZER

MDR report key: 3480446 · Received November 8, 2013

Report

Report Number
1054871-2013-00091
Event Type
Injury
Date Received
November 8, 2013
Date of Event
October 15, 2013
Report Date
November 11, 2013
Manufacturer
HEALTH & LIFE (SUZHOU) CO., LTD.
Product Code
CCQ
Product Problem
Yes
Report Source
Distributor report
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE PT CONTACTED NEPHRON PHARMACEUTICALS CORP REGARDING A PRODUCT COMPLAINT ASSOCIATED WITH THE EZ BREATH ATOMIZER ON (B)(6) 2013. THE PT REPORTED THAT THE ATOMIZER FAILED TO PRODUCE AN AEROSOL MIST. IN ADDITION, THE PT ADDED THAT SHE REQUIRED MEDICAL TREATMENT AT THE HOSPITAL. MULTIPLE ATTEMPTS WERE MADE TO REACH THE CUSTOMER VIA TELEPHONE AND MAIL UNSUCCESSFULLY. THE INVESTIGATED PRODUCT FOR THE ENCLOSED MEDICAL DEVICE REPORT IS LISTED AMONG THE IMPLICATED LOTS FOR A NATIONWIDE RECALL INITIATED BY THE MFR (B)(4)., ON (B)(4) 2013. THE CLASS 1 RECALL (Z-1371-2013, Z-1372-2013, Z-1373-2013) WAS INITIATED AFTER NEPHRON PHARMACEUTICALS CORP, THE IMPORTER, BECAME AWARE OF AN INCREASING NUMBER OF COMPLAINTS ASSOCIATED WITH THE POSSIBILITY OF A QUARTER-INCH WASHER BECOMING DISLODGED FROM THE EZ BREATH ATOMIZER. THE IMPACTED ATOMIZER SERIAL NUMBER RANGES ARE: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579048 EZ BREATH ATOMIZER ATOMIZER CCQ HEALTH & LIFE (SUZHOU) CO., LTD. EZ-100 121001

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization| O UNIDENTIFIED